Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Beckman DM5XL: CAL ADJ /5500 INSTRUMENT: Beckman DM10XL: CAL ADJ /5500 INSTRUMENT: Beckman DM15XL: CAL ADJ /5500 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 67 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # dm5xlv.bmp # dm10xlv.bmp # dm15xlv.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 minutes 1.004 DISP Temperature: 25 +/- 2degC 1.004 DISP Humidity: <70% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD DISASSEMBLY INSTRUCTIONS 1.007 DISP 1. Remove the back case screws, and carefully 1.007 DISP [32] pry up the back case. 1.007 DISP 1.007 DISP 2. Set the FUNCTION/RANGE switch to 200mV DC. 1.008 HEAD DC VOLTAGE 1.009 JMPT 1.012 BECKMAN DM5XL: CAL ADJ /5500 1.010 JMPT 1.014 BECKMAN DM10XL: CAL ADJ /5500 1.011 JMP 1.016 1.012 PIC dm5xlv 1.013 JMP 1.017 1.014 PIC dm10xlv 1.015 JMP 1.017 1.016 PIC dm15xlv 1.017 5500 190.0mV S 2W 1.018 DISP ADJUST VR1 until display reads 190.0mV +/- 1digit. 1.019 5500 * S 1.020 DISP Disconnect ALL connections. 1.021 HEAD ASSEMBLY INSTRUCTIONS 1.022 DISP Reassemble the back case with the two screws. 1.023 END